Материал: part16

Внимание! Если размещение файла нарушает Ваши авторские права, то обязательно сообщите нам

 

 

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 201​

 

NL​

Rel with Parent​

VT​

Concept Name​

VM​

Req​ Condition​

Value Set​

 

 

 

 

 

 

 

Type​

Constraint​

5​

>​

HAS ACQ​

INCLUDE​

DTID 3107 “Device Used”​

1-n​

U​

 

 

 

CONTEXT​

 

 

 

 

 

 

6​

>​

CONTAINS​

INCLUDE​

DTID 3253 “IVUS​

1​

MC​ One or both of​

 

 

 

 

 

 

Measurements”​

 

rows 6 & 7​

 

 

 

 

 

 

 

 

must be​

 

 

 

 

 

 

 

 

present​

 

7​

>​

CONTAINS​

INCLUDE​

DTID 3254 “IVUS​

1​

MC​ One or both of​

 

 

 

 

 

 

Qualitative Assessments”​

 

rows 6 & 7​

 

 

 

 

 

 

 

 

must be​

 

 

 

 

 

 

 

 

present​

 

8​

>​

CONTAINS​

INCLUDE​

DTID 3819 “Common​

1-n​

U​

 

 

 

 

 

 

Findings”​

 

 

 

Content Item Descriptions​

 

 

 

 

 

 

Row 2​

Lesion Identifier is specified as a numeric text string in order to facilitate trans-coding to DICOM Attribute​

 

 

(0018,3105) Lesion Number and to formats for outcomes registries, such as the ACC National Cardiovascular​

 

 

Data Registry™.​

 

 

 

 

 

 

 

Note​

 

 

 

 

 

 

 

 

Also see TID 3105 “Lesion Identification and Properties”.​

 

 

Row 3​

Finding site may span multiple segments with the proximal and distal extent specified by separate items. These​

 

 

may not be totally contained with the segment specified at the Vessel level.​

 

TID 3253 IVUS Measurements​

TheIVUSmeasurementsTemplategroupstogethersimpledistance,areaandanglemeasurements,alongwithderivedmeasurements​ that made during an IVUS procedure. Refer to the " ACC Clinical Expert Consensus Document on Standards for Acquisition, meas-​ urement and Reporting of Intravascular Ultrasound Studies (IVUS) "for more information.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

 

Table TID 3253. IVUS Measurements​

NL​ Rel​

VT​

Concept Name​

VM​

Req​Condition​

Value Set Constraint​

with​

 

 

 

Type​

 

 

Parent​

 

 

 

 

 

1​

INCLUDE​

DTID 300​

1-n​

U​

$Measurement = DCID 3481 “IVUS Distance​

 

 

“Measurement”​

 

 

Measurements”​

$Units = EV (mm, UCUM, "mm")​

$Derivation = DCID 3488 “Min/Max/Mean”​

$TargetSite = BCID 3486 “Vascular​

Measurement Sites”​

- Standard -​

Page 202​

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

NL​ Rel​

VT​

Concept Name​

VM​

Req​Condition​

Value Set Constraint​

with​

 

 

 

Type​

 

 

Parent​

 

 

 

 

 

2​

INCLUDE​

DTID 300​

1-n​

U​

$Measurement = DCID 3482 “IVUS Area​

 

 

“Measurement”​

 

 

Measurements”​

 

 

 

 

 

$Units = EV (mm2, UCUM, "mm2")​

 

 

 

 

 

$Derivation = DCID 3488 “Min/Max/Mean”​

 

 

 

 

 

$TargetSite = BCID 3486 “Vascular​

 

 

 

 

 

Measurement Sites”​

3​

INCLUDE​

DTID 300​

1-n​

U​

$Measurement = DCID 3483 “IVUS​

 

 

“Measurement”​

 

 

Longitudinal Measurements”​

 

 

 

 

 

$Units = EV (mm, UCUM, "mm")​

4​

INCLUDE​

DTID 300​

1-n​

U​

$Measurement = EV (122355, DCM, "Arc of​

 

 

“Measurement”​

 

 

Calcium")​

 

 

 

 

 

$Units = EV (deg, UCUM, "degrees")​

 

 

 

 

 

$TargetSite = BCID 3486 “Vascular​

 

 

 

 

 

Measurement Sites”​

5​

INCLUDE​

DTID 300​

1​

U​

$Measurement = EV (408714007, SCT,​

 

 

“Measurement”​

 

 

"Lumen Area Stenosis")​

 

 

 

 

 

$Units = EV (%, UCUM, "%")​

6​

INCLUDE​

DTID 300​

1​

U​

$Measurement = EV (122354, DCM, "Plaque​

 

 

“Measurement”​

 

 

Burden")​

 

 

 

 

 

 

$Units = EV (%, UCUM, "%")​

 

 

 

 

 

$TargetSite = BCID 3486 “Vascular​

 

 

 

 

 

Measurement Sites”​

7​

INCLUDE​

DTID 300​

1-n​

U​

$Measurement = DCID 3484 “IVUS Indices​

 

 

“Measurement”​

 

 

and Ratios”​

 

 

 

 

 

$Units = EV ({ratio}, UCUM, "ratio")​

 

 

 

 

 

$TargetSite = BCID 3486 “Vascular​

 

 

 

 

 

Measurement Sites”​

8​

INCLUDE​

DTID 3255 “IVUS​

1-n​

U​

 

 

 

 

Volume​

 

 

 

 

 

 

Measurement”​

 

 

 

 

9​

INCLUDE​

DTID 300​

1​

U​

$Measurement = EV (122339, DCM, "Stent​

 

 

“Measurement”​

 

 

Volume Obstruction")​

$Units = EV (%, UCUM, "%")​

TID 3254 IVUS Qualitative Assessments​

The IVUS Qualitative Assessments Template groups together the qualitative properties of a lesion that are observed during an IVUS​ procedure. Refer to the "ACC Clinical Expert Consensus Document on Standards for Acquisition, measurement and Reporting of​ Intravascular Ultrasound Studies (IVUS) " for more information.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

- Standard -​

DICOM PS3.16 2020a - Content Mapping Resource​ Page 203​

Table TID 3254. IVUS Qualitative Assessments​

 

NL​ Rel with​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

Parent​

 

 

 

Type​

 

1​

 

CODE​

EV (122133, DCM, "Lesion​

1-n​

U​

DCID 3491 “IVUS Lesion​

 

 

 

Morphology")​

 

 

Morphologies”​

2​

 

CODE​

EV (121071, DCM, "Finding")​

1-n​

U​

DCID 3494 “IVUS Non​

 

 

 

 

 

 

Morphological Findings”​

3​

>​

INCLUDE​

DTID 1350 “Negation Modifier,​

1​

U​

 

 

 

 

Presence of Finding”​

 

 

 

4​

 

CODE​

EV (121071, DCM, "Finding")​

1​

U​

EV (710864009, SCT,​

 

 

 

 

 

 

"Arterial Dissection")​

5​

>​ HAS​

CODE​

EV (122387, DCM, "Dissection​

1​

U​

DCID 3492 “Vascular​

 

CONCEPT​

Classification")​

 

 

Dissection Classifications”​

 

MOD​

 

 

 

 

 

6​

>​

INCLUDE​

DTID 1350 “Negation Modifier,​

1​

U​

 

 

 

 

Presence of Finding”​

 

 

 

7​

 

CODE​

EV (122391, DCM, "Relative​

1​

U​

DCID 3493 “IVUS Relative​

 

 

 

Stenosis Severity")​

 

 

Stenosis Severities”​

8​

 

CODE​

EV (108, NCDR [2.0b],​

1​

U​

DCID 3750 “Previously​

 

 

 

"Previously Dilated Lesion ")​

 

 

Dilated Lesion”​

9​

 

CODE​

EV (121071, DCM, "Finding")​

1​

U​

EV (122393, DCM,​

 

 

 

 

 

 

"Restenotic Lesion")​

10​

>​

INCLUDE​

DTID 1350 “Negation Modifier,​

1​

U​

 

 

 

 

Presence of Finding”​

 

 

 

11​

 

CODE​

EV(111009,DCM,"Calcification​

1​

U​

DCID 3489 “Calcium​

 

 

 

Type")​

 

 

Distribution”​

TID 3255 IVUS Volume Measurement​

The IVUS Volume Measurement Template contains information describing an IVUS Volumetric measurement​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

 

Table TID 3255. IVUS Volume Measurement​

 

NL​ Rel with Parent​ VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

 

 

 

 

 

Type​

 

1​

 

 

INCLUDE​DTID 300 “Measurement”​ 1​

M​

$Measurement = DCID 3485​

 

 

 

 

 

 

 

“IVUS Volume Measurements”​

 

 

 

 

 

 

 

$Units = EV (mm3, UCUM,​

 

 

 

 

 

 

 

"mm3")​

 

 

 

 

 

 

 

$TargetSite = BCID 3487​

 

 

 

 

 

 

 

“Intravascular Volumetric​

 

 

 

 

 

 

 

Regions”​

2​

>​

HAS​

NUM​

EV (122336, DCM,​

1​

U​

UNITS=DT(mm,UCUM,"mm")​

 

 

PROPERTIES​

 

"Vascular Volume​

 

 

 

 

 

 

 

measurement length")​

 

 

 

3​

>​

HAS​

NUM​

EV (122337, DCM,​

1​

U​

UNITS=DT(mm,UCUM,"mm")​

 

 

PROPERTIES​

 

"Relative position")​

 

 

 

- Standard -​

Page 204​

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

NL​ Rel with Parent​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

 

 

 

 

Type​

 

4​ >>​

HAS CONCEPT​CODE​

EV (122340, DCM,​

1​

M​

DCID 3496 “IVUS Fiducial​

 

MOD​

 

"Fiducial feature")​

 

 

Points”​

Stress Testing Report Templates ​

TID 3300 Stress Testing Report​

The Stress Testing Report Template is the root structure for the representation of measurements and findings of a stress testing​ procedure.​

Type:​

Extensible​

Order:​

Significant​

Root:​

Yes​

Table TID 3300. Stress Testing Report​

 

NL​

Rel with Parent​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set​

 

 

 

 

 

 

 

Type​

Constraint​

1​

 

 

CONTAINER​ DT(18752-6,LN,"StressTesting​

1​

M​

Root node​

 

 

 

 

 

Report")​

 

 

 

2​

>​

HAS CONCEPT​ CODE​

EV (121058, DCM, "Procedure​

1​

M​

BCID 3200​

 

 

MOD​

 

 

reported")​

 

 

“Stress Test​

 

 

 

 

 

 

 

 

Procedure”​

3​

>​

HAS CONCEPT​ INCLUDE​

DTID1204“LanguageofContent​

1​

M​

 

 

 

MOD​

 

 

Item and Descendants”​

 

 

 

4​

>​

HAS OBS​

INCLUDE​

DTID 1002 “Observer Context”​

1-n​

M​

 

 

 

CONTEXT​

 

 

 

 

 

 

5​

>​

CONTAINS​

CONTAINER​ EV(18785-6,LN,"Indicationsfor​

1​

U​

 

 

 

 

 

 

Procedure")​

 

 

 

6​

>>​

CONTAINS​

CODE​

EV (121071, DCM, "Finding")​

1-n​

U​

DCID 3201​

 

 

 

 

 

 

 

 

“Indications​

 

 

 

 

 

 

 

 

for Stress​

 

 

 

 

 

 

 

 

Test”​

7​

>>​

CONTAINS​

TEXT​

EV (121071, DCM, "Finding")​

1​

U​

 

8​

>​

CONTAINS​

INCLUDE​

DTID 3602 “Cardiovascular​

1​

M​

 

 

 

 

 

 

Patient Characteristics”​

 

 

 

9​

>​

CONTAINS​

INCLUDE​

DTID 3802 “Cardiovascular​

1​

U​

 

 

 

 

 

 

Patient History”​

 

 

 

10​

>​

CONTAINS​

INCLUDE​

DTID 3301 “Stress Test​

1​

M​

 

 

 

 

 

 

Procedure Description”​

 

 

 

11​

>​

CONTAINS​

INCLUDE​

DTID 3303 “Stress Test Phase​

1-n​

M​

 

 

 

 

 

 

Data”​

 

 

 

12​

>​

CONTAINS​

INCLUDE​

DTID 3311 “Stress Test​

1​

U​

 

 

 

 

 

 

Summary”​

 

 

 

13​

>​

CONTAINS​

INCLUDE​

DTID 3318 “Comparison to Prior​

1​

U​

 

 

 

 

 

 

Stress Exam”​

 

 

 

14​

>​

CONTAINS​

INCLUDE​

DTID 3320 “Conclusions and​

1​

U​

 

 

 

 

 

 

Recommendations”​

 

 

 

TID 3301 Stress Test Procedure Description​

Type:​ Extensible​

- Standard -​

 

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 205​

Order:​

 

Significant​

 

 

 

Root:​

 

No​

 

 

 

 

 

 

 

Table TID 3301. Stress Test Procedure Description​

 

 

NL​

Rel with​

VT​

Concept Name​

VM​

Req​ Condition​

Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

1​

 

 

CONTAINER​DT (55111-9, LN, "Current​

1​

M​

 

 

 

 

 

Procedure Descriptions")​

 

 

 

2​

>​

CONTAINS​

CODE​

DT (109056, DCM, "Stress​

1​

U​

BCID 3261 “Stress​

 

 

 

 

Protocol")​

 

 

Protocols”​

3​

>​

CONTAINS​

TEXT​

DT (109056, DCM, "Stress​

1​

U​

 

 

 

 

 

Protocol")​

 

 

 

4​

>​

CONTAINS​

CODE​

DT (10:11345, MDC, "Lead​ 1​

U​

BCID 3263 “Electrode​

 

 

 

 

System")​

 

 

Placement Values”​

5​

>​

CONTAINS​

CODE​

DT (111045004, SCT,​

1​

U​

BCID 3203 “Exerciser​

 

 

 

 

"Exerciser Device")​

 

 

Device”​

6​

>​

CONTAINS​

CODE​

DT (246489000, SCT,​

1​

MC​ IFF​

BCID 3204 “Stress​

 

 

 

 

"Pharmacological Stress​

 

Pharmacological​Agents”​

 

 

 

 

Agent")​

 

Stress used​

 

7​

>​

CONTAINS​

CONTAINER​EV (122700, DCM,​

1​

MC​ IFF​

 

 

 

 

 

"Indications for​

 

Pharmacological​

 

 

 

 

 

Pharmacological Stress")​

 

Stress used​

 

8​

>>​

CONTAINS​

CODE​

EV (121071, DCM,​

1-n​

U​

DCID3205“Indications​

 

 

 

 

"Finding")​

 

 

for Pharmacological​

 

 

 

 

 

 

 

Stress Test”​

9​

>​

CONTAINS​

CODE​

DT (363679005, SCT,​

1​

MC​ IFF imaging used​DCID 3206​

 

 

 

 

"Imaging procedure")​

 

in procedure​

“Non-invasive Cardiac​

 

 

 

 

 

 

 

Imaging Procedures”​

10​

>​

CONTAINS​

CODE​

DT (125203, DCM,​

1​

UC​ IFF Nuclear​

DCID 3110 “Nuclear​

 

 

 

 

"Acquisition Protocol")​

 

imaging​

Cardiology Protocols”​

11​

>​

CONTAINS​

CODE​

DT (125203, DCM,​

1​

UC​ IFF PET imaging​DCID 3106 “PET​

 

 

 

 

"Acquisition Protocol")​

 

 

Cardiology Protocols”​

12​

>​

CONTAINS​

TEXT​

DT (121141, DCM, "Image​

1​

UC​ IFF Nuclear or​

STATIC,DYNAMIC,or​

 

 

 

 

Type")​

 

PET imaging​

GATED.​

13​

>​

CONTAINS​

CODE​

DT (RID11248, RADLEX,​

1​

U​

DCID 3104 “Cardiac​

 

 

 

 

"Cardiac Gating")​

 

 

Synchronization​

 

 

 

 

 

 

 

Technique”​

14​

>​

CONTAINS​

CODE​

DT (125203, DCM,​

1​

UC​ IFF Contrast​

DT (433231002, SCT,​

 

 

 

 

"Acquisition Protocol")​

 

echocardiography​"Contrast​

 

 

 

 

 

 

 

echocardiography")​

15​

>​

CONTAINS​

CODE​

DT (113743, DCM, "Patient​ 1​

U​

DCID 19 “Patient​

 

 

 

 

Orientation")​

 

 

Orientation”​

16​

>>​

HAS​

CODE​

EV (113744, DCM, "Patient​ 1​

U​

DCID 20 “Patient​

 

 

CONCEPT​

 

Orientation Modifier")​

 

 

Orientation Modifier”​

 

 

MOD​

 

 

 

 

 

17​

>​

CONTAINS​

TEXT​

DT (121065, DCM,​

1​

U​

 

 

 

 

 

"Procedure Description")​

 

 

 

18​

>​

CONTAINS​

DATETIME​ DT (122701, DCM,​

1​

U​

 

 

 

 

 

"Procedure Time Base")​

 

 

 

Content Item Descriptions​

- Standard -​

Источник: https://studfile.net/preview/14585769/