DICOM PS3.16 2020a - Content Mapping Resource Page 171
Table TID 3108. Intervention
|
NL |
Rel with Parent VT |
Concept Name |
VM |
Req ConditionValue Set Constraint |
||
|
|
|
|
|
|
Type |
|
1 |
|
|
CODE |
EV (122090, DCM, |
1 |
M |
DCID 3412 |
|
|
|
|
"Intervention Action") |
|
|
“Intervention Actions |
|
|
|
|
|
|
|
and Status” |
2 |
> |
HAS |
CODE |
EV (363704007, SCT, |
1 |
M |
DCID 3604 “Arterial |
|
|
PROPERTIES |
|
"Procedure site") |
|
|
Lesion Locations” |
3 |
>> |
HAS CONCEPT CODE |
EV (106233006, SCT, |
1 |
U |
DCID 3019 |
|
|
|
MOD |
|
"Topographical modifier") |
|
|
“Cardiovascular |
|
|
|
|
|
|
|
Anatomic Location |
|
|
|
|
|
|
|
Modifiers” |
4 |
> |
HAS |
TEXT |
EV (121154, DCM, |
1 |
M |
Up to 3 numeric |
|
|
PROPERTIES |
|
"Intervention attempt |
|
|
characters |
|
|
|
|
identifier") |
|
|
|
5 |
> |
HAS |
CODE |
EV (116682006, SCT, "Uses |
1-n |
U |
BCID 3411 |
|
|
PROPERTIES |
|
Equipment") |
|
|
“IntracoronaryDevices” |
6 |
>> |
HAS CONCEPT CODE |
EV (122111, DCM, "Primary |
1 |
MC IF DeviceDCID 230 “Yes-No” |
||
|
|
MOD |
|
Intervention Device") |
|
|
is Primary |
|
|
|
|
|
|
|
for this |
|
|
|
|
|
|
|
Lesion |
7 |
> |
HAS |
NUM |
DCID 3425 “Intervention |
1-n |
U |
|
|
|
PROPERTIES |
|
Parameters” |
|
|
|
8 |
> |
HAS |
IMAGE |
BCID 7003 “Diagnostic |
1 |
U |
|
|
|
PROPERTIES |
|
Imaging Report Purposes of |
|
|
|
|
|
|
|
Reference” |
|
|
|
9 |
> |
|
INCLUDE DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
|
|
Qualifiers” |
|
|
|
Content Item Descriptions |
|
|
|
|
|
||
Row 4 |
Intervention attempt Identifier is specified as a numeric text string, and shall be treated as the ordinal of the |
||||||
|
|
recorded attempted intervention within this procedure (i.e., "1" for the first attempted intervention, "2" for the |
|||||
|
|
second, etc.). |
|
|
|
|
|
TID 3109 Measurements
The Measurements Template allows recording of significant measurements, such as vital signs, laboratory results, hemodynamic measurements, or quantitative analysis measurements. These measurements are often quoted from another source, which would be documented in the included TID 3010 “Log Entry Qualifiers”.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
|
Table TID 3109. Measurements |
|
NL |
RelwithParent |
VT |
Concept Name |
VM |
Req TypeConditionValue Set |
|
|
|
|
|
|
|
|
Constraint |
1 |
|
|
NUM |
|
1 |
U |
|
2 |
> |
|
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
|
|
Qualifiers” |
|
|
3 |
> |
HAS |
INCLUDE |
DTID 310 “Measurement |
1 |
U |
|
|
|
PROPERTIES |
|
Properties” |
|
|
|
- Standard -
Page 172 |
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
|||
|
NL |
RelwithParent |
VT |
Concept Name |
VM |
Req TypeConditionValue Set |
|
|
|
|
|
|
Constraint |
4 |
|
CODE |
|
1 |
U |
|
5 |
> |
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
|
Qualifiers” |
|
|
TID 3110 Impressions or Findings
The Impressions or Findings Template allows the recording of unconfirmed (provisional) impressions or findings noted during the procedure. It is not intended to convey the Cath Lab Clinical Report (the formal report from the performing physician), although it may be used (like any Procedure Log entry) for the subsequent construction of the Cath Lab Clinical Report.
A finding that is supported by a specific image frame may reference that image in the INFERRED FROM / IMAGE row of the included
TID 3010 “Log Entry Qualifiers” Template.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
|
Table TID 3110. Impressions or Findings |
|
NL Rel with Parent VT |
Concept Name |
VM |
Req Condition Value Set |
|||
|
|
|
|
|
|
Type |
Constraint |
1 |
|
|
CODE |
EV (121071, DCM, "Finding") |
1 |
U |
BCID 3728 “Cath |
|
|
|
|
|
|
|
Findings” |
2 |
> |
HAS |
CODE |
EV (246112005, SCT, |
1 |
U |
DCID3716“Severity” |
|
|
PROPERTIES |
|
"Severity") |
|
|
|
3 |
> |
HAS |
CODE |
EV (363698007, SCT, |
1 |
U |
|
|
|
PROPERTIES |
|
"Finding Site") |
|
|
|
4 |
>> |
HAS CONCEPTCODE |
EV (106233006, SCT, |
1 |
U |
DCID 3019 |
|
|
|
MOD |
|
"Topographical modifier") |
|
|
“Cardiovascular |
|
|
|
|
|
|
|
Anatomic Location |
|
|
|
|
|
|
|
Modifiers” |
5 |
> |
|
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
|
Qualifiers” |
|
|
|
6 |
|
|
TEXT |
BCID 3419 “Findings Titles” |
1 |
U |
|
7 |
> |
|
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
|
Qualifiers” |
|
|
|
Content Item Descriptions |
|
|
|
|
|
||
Row 3 |
Finding Site has no Baseline Context ID specified. Typically terms would be drawn from coronary segments, |
||||||
|
|
other arterial segments, myocardial segments, etc. |
|
|
|
||
TID 3111 Percutaneous Entry
The Percutaneous Entry Template allows recording of the opening or closing of invasive access ports.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
- Standard -
DICOM PS3.16 2020a - Content Mapping Resource Page 173
Table TID 3111. Percutaneous Entry
|
NL RelwithParent VT |
Concept Name |
VM Req TypeConditionValue Set Constraint |
|||
1 |
|
CODE |
EV (121156, DCM, |
1 |
M |
DCID 3403 |
|
|
|
"PercutaneousEntryAction") |
|
|
“Percutaneous Entry” |
2 |
> |
HASCONCEPTCODE |
EV (272741003, SCT, |
1 |
U |
DCID 244 “Laterality” |
|
|
MOD |
"Laterality") |
|
|
|
3 |
> |
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
Qualifiers” |
|
|
|
TID 3112 Specimen Obtained
The Specimen Obtained Template allows recording of obtaining a specimen, and the identifiers for that specimen. This is particularly designedforbloodsamplesthatwillbeanalyzedforbloodoxygen-relatedmeasurements.Theanalysisofthesamplemayberecorded in one or more log entries using TID 3109 “Measurements”, or in a separate Structured Report SOP Instance referenced by a log entry using TID 3103 “Referenced Object”.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
Table TID 3112. Specimen Obtained
|
NL Rel with Parent VT |
Concept Name |
VM |
Req |
Condition Value Set Constraint |
|||
|
|
|
|
|
|
Type |
|
|
1 |
|
|
CODE |
EV (121123, DCM, "Patient |
1 |
M |
|
DCID 3515 “Specimen |
|
|
|
|
Status or Event") |
|
|
|
Collection” |
2 |
> |
HAS ACQ |
CODE |
EV (371439000, SCT, |
1 |
UC |
IFF |
DCID 3520 “Blood |
|
|
CONTEXT |
|
"Specimen Type") |
|
|
specimen isSource Type” |
|
|
|
|
|
|
|
|
bloodsample |
|
3 |
> |
HAS ACQ |
CODE |
EV (363704007, SCT, |
1 |
U |
|
BCID 3630 |
|
|
CONTEXT |
|
"Procedure site") |
|
|
|
“Cardiovascular |
|
|
|
|
|
|
|
|
Anatomic Locations” |
4 |
> |
HAS |
INCLUDE DTID 1009 “Subject |
1 |
U |
|
|
|
|
|
PROPERTIES |
|
Context, Specimen” |
|
|
|
|
TID 3113 Patient Support
The Patient Support Template allows recording of the use of various support technologies, including oxygen, ventilation, pacing, etc.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
|
Table TID 3113. Patient Support |
NL |
Rel with |
VT |
Concept Name |
VM |
Req Condition |
Value Set |
|
Parent |
|
|
|
Type |
Constraint |
1 |
|
CODE |
DCID 3530 “Oxygen |
1 |
U |
DCID 3531 “Oxygen |
|
|
|
Administration Actions” |
|
|
Administration” |
2 > |
HAS |
NUM |
EV (121160, DCM, "Oxygen |
1 |
MC IFF Row 1 |
UNITS = DT (l/min, |
|
PROPERTIES |
Administration Rate") |
|
Concept is |
UCUM, "l/min") |
|
|
|
|
|
|
(121161, DCM, |
|
|
|
|
|
|
"Begin Oxygen |
|
|
|
|
|
|
Administration") |
|
- Standard -
Page 174 |
DICOM PS3.16 2020a - Content Mapping Resource |
|
||||
|
NL Rel with |
VT |
Concept Name |
VM |
Req Condition |
Value Set |
|
Parent |
|
|
|
Type |
Constraint |
3 |
> |
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
Qualifiers” |
|
|
|
4 |
|
CODE |
DCID 3550 “Circulatory |
1 |
U |
DCID 3553 |
|
|
|
Support Actions” |
|
|
“CirculatorySupport” |
5 |
> |
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
Qualifiers” |
|
|
|
6 |
|
CODE |
DCID 3551 “Ventilation |
1 |
U |
DCID 3554 |
|
|
|
Actions” |
|
|
“Ventilation” |
7 |
> |
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
Qualifiers” |
|
|
|
8 |
|
CODE |
DCID 3552 “Pacing Actions” |
1 |
U |
DCID 3555 “Pacing” |
9 |
> |
INCLUDE |
DTID 3010 “Log Entry |
1 |
U |
|
|
|
|
Qualifiers” |
|
|
|
TID 3114 Patient Assessment
ThePatientAssessmentTemplateallowsrecordingoftheassessmentofthepatient'scardiovascular,neurological,and/orrespiratory condition. A particular use of this Template is for "vital signs", which are a specific subset of mandatory patient assessment measure- ments.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
|
Table TID 3114. Patient Assessment |
|
NL Rel with |
VT |
Concept Name |
VM Req |
Condition |
Value Set Constraint |
|
|
|
Parent |
|
|
Type |
|
|
1 |
|
|
CODE |
EV (121123, DCM, |
1 M |
|
DT (121165, DCM, "Patient |
|
|
|
|
"Patient Status or |
|
|
Assessment Performed") |
|
|
|
|
Event") |
|
|
DT (61746007, SCT, |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
"Observation of Vital Signs") |
2 |
> |
HAS |
INCLUDEDTID 300 |
1 MCIF Row 1 value = |
$Measurement = EV |
||
|
|
PROPERTIES |
“Measurement” |
|
(61746007, SCT, |
(271649006, SCT, "Systolic |
|
|
|
|
|
|
|
"Observation of Vitalblood pressure") |
|
|
|
|
|
|
|
Signs") |
$Units = DCID 3500 “Pressure |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Units” |
|
|
|
|
|
|
|
$Method = BCID 3560 “Blood |
|
|
|
|
|
|
|
Pressure Methods” |
3 |
> |
HAS |
INCLUDEDTID 300 |
1 MCIF Row 1 value = |
$Measurement = EV |
||
|
|
PROPERTIES |
“Measurement” |
|
(61746007, SCT, |
(271650006, SCT, "Diastolic |
|
|
|
|
|
|
|
"Observation of Vitalblood pressure") |
|
|
|
|
|
|
|
Signs") |
$Units = DCID 3500 “Pressure |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Units” |
4 |
> |
HAS |
INCLUDEDTID 300 |
1 MCIF Row 1 value = |
$Measurement = EV (8867-4, |
||
|
|
PROPERTIES |
“Measurement” |
|
(61746007, SCT, |
LN, "Heart rate") |
|
|
|
|
|
|
|
"Observation of Vital |
|
|
|
|
|
|
|
Signs") |
$Units=EV({H.B.}/min,UCUM, |
|
|
|
|
|
|
|
"BPM") |
- Standard -
|
|
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 175 |
|||
|
NL Rel with |
VT |
Concept Name |
VM Req |
Condition |
Value Set Constraint |
||
|
|
Parent |
|
|
|
Type |
|
|
5 |
> |
HAS |
INCLUDEDTID 300 |
1 MCIF Row 1 value = |
$Measurement = EV (8310-5, |
|||
|
|
PROPERTIES |
“Measurement” |
|
|
(61746007, SCT, |
LN, "Body temperature") |
|
|
|
|
|
|
|
|
"Observation of Vital |
|
|
|
|
|
|
|
|
Signs") |
$Units = EV (Cel, UCUM, "C") |
6 |
> |
HAS |
INCLUDEDTID 300 |
1 MCIF Row 1 value = |
$Measurement = DCID 3526 |
|||
|
|
PROPERTIES |
“Measurement” |
|
|
(61746007, SCT, |
“Blood Gas Saturation” |
|
|
|
|
|
|
|
|
"Observation of Vital |
|
|
|
|
|
|
|
|
Signs") |
$Units = EV (%, UCUM, "%") |
7 |
> |
HAS |
INCLUDEDTID 300 |
1 MCIF Row 1 value = |
$Measurement=EV(86290005, |
|||
|
|
PROPERTIES |
“Measurement” |
|
|
(61746007, SCT, |
SCT, "Respiration rate") |
|
|
|
|
|
|
|
|
"Observation of Vital |
|
|
|
|
|
|
|
|
Signs") |
$Units = EV (/min, UCUM, |
|
|
|
|
|
|
|
|
"breaths/min") |
8 |
> |
HAS |
INCLUDEDTID 300 |
1-n MCIF Row 1 value = |
$Measurement = EV (122195, |
|||
|
|
PROPERTIES |
“Measurement” |
|
|
(61746007, SCT, |
DCM, "Pulse Strength") |
|
|
|
|
|
|
|
|
"Observation of Vital |
|
|
|
|
|
|
|
|
Signs") |
$Method = BCID 3442 |
|
|
|
|
|
|
|
|
“Peripheral Pulse Methods” |
|
|
|
|
|
|
|
|
$TargetSite = BCID 3440 |
|
|
|
|
|
|
|
|
“Peripheral Pulse Locations” |
|
|
|
|
|
|
|
|
$Units = DT ({0:4}, UCUM, |
|
|
|
|
|
|
|
|
"range 0:4") |
9 |
> |
HAS |
INCLUDEDTID 300 |
1 MCIF Row 1 value = |
$Measurement = EV |
|||
|
|
PROPERTIES |
“Measurement” |
|
|
(61746007, SCT, |
(225908003,SCT,"PainScore") |
|
|
|
|
|
|
|
|
"Observation of Vital |
|
|
|
|
|
|
|
|
Signs") |
$Units = DT ({1:10}, UCUM, |
|
|
|
|
|
|
|
|
"range 1:10") |
10 |
> |
HAS |
CODE |
DT (8884-9, LN, |
1 |
U |
|
BCID 3415 “Cardiac Rhythms” |
|
|
PROPERTIES |
"Cardiac Rhythm") |
|
|
|
|
|
11 |
> |
HAS |
CODE |
DT (9304-7, LN, |
1 |
U |
|
BCID 3416 “Respiration |
|
|
PROPERTIES |
"Respiration |
|
|
|
Rhythms” |
|
|
|
|
|
Rhythm") |
|
|
|
|
12 |
> |
HAS |
CODE |
DT (364062005, |
1 |
U |
|
BCID3448“AirwayAssessment” |
|
|
PROPERTIES |
SCT, "Respiration |
|
|
|
|
|
|
|
|
|
Assessment") |
|
|
|
|
13 |
> |
HAS |
CODE |
DT (364528001, |
1-n |
U |
|
BCID 3446 “Skin Condition” |
|
|
PROPERTIES |
SCT, "Skin |
|
|
|
|
|
|
|
|
|
condition") |
|
|
|
|
14 > |
HAS |
CODE |
DT (363871006, |
1 |
U |
|
|
|
|
|
PROPERTIES |
SCT, "Patient mental |
|
|
|
||
|
|
|
|
state assessment") |
|
|
|
|
15 |
> |
HAS |
TEXT |
BCID 3441 “Patient 1-n |
U |
|
|
|
|
|
PROPERTIES |
Assessments” |
|
|
|
|
|
Content Item Descriptions |
|
|
|
|
|
|||
Row 8 |
Pulse Strength allows the assessment of the patient's pulse at multiple locations using the Topographical |
|||||||
|
|
concept modifier. It may also be used for a single pulse strength measurement from an unspecified location, |
||||||
|
|
as is typical of vital signs assessments. |
|
|
|
|||
Row 16 |
Allows free text description of patient assessments that are not expressible by coded entries of Rows 10 to |
|||||||
|
|
14. |
|
|
|
|
|
|
- Standard -