DICOM PS3.16 2020a - Content Mapping Resource​

Page 221​

Row 4​

Procedure Action ID allows linkage between the hemodynamic measurements recorded in this Template and​

 

a procedure step or phase recorded in the Procedure Log, e.g., using TID 3100 “Procedure Action”.​

TID 3504 Arterial Pressure Measurement​

The Arterial Pressure Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic​ site, and containing systolic, diastolic, and mean measurements. This implies that the name of the measurement is effectively post-​ coordinated from the measurement name, the Hemodynamic Measurement Group container (procedure phase) name, and the ac-​ quisition context (finding site) value.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

 

Table TID 3504. Arterial Pressure Measurement​

 

NL​

Rel with​

VT​

Concept Name​

VM​

Req​ Condition​

Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

 

1​

 

 

CONTAINER​EV (73002000, SCT,​

1​

M​

 

 

 

 

 

 

"Arterial pressure​

 

 

 

 

 

 

 

 

measurements")​

 

 

 

 

2​

>​

 

INCLUDE​

DTID 3530​

1​

M​

$LocationName = EV (363698007,​

 

 

 

 

“Hemodynamic​

 

 

SCT, "Finding Site")​

 

 

 

 

Acquisition Context”​

 

 

$LocationValue = DCID 3606​

 

 

 

 

 

 

 

 

 

 

 

 

 

 

“Arterial Source Locations”​

3​

>​

CONTAINS​INCLUDE​

DTID300“Measurement”​ 1​

M​

$Measurement = EV (8480-6, LN,​

 

 

 

 

 

 

 

"Intravascular arterial Systolic​

 

 

 

 

 

 

 

pressure")​

 

 

 

 

 

 

 

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

Units”​

4​

>​

CONTAINS​INCLUDE​

DTID300“Measurement”​ 1​

M​

$Measurement = EV (8462-4, LN,​

 

 

 

 

 

 

 

"Intravascular arterial Diastolic​

 

 

 

 

 

 

 

pressure")​

 

 

 

 

 

 

 

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

Units”​

5​

>​

CONTAINS​INCLUDE​

DTID300“Measurement”​ 1​

M​

$Measurement = EV (8478-0, LN,​

 

 

 

 

 

 

 

"Intravascular arterial mean​

 

 

 

 

 

 

 

pressure")​

 

 

 

 

 

 

 

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

Units”​

6​

>​

CONTAINS​INCLUDE​

DTID 3550 “Pressure​

1-n​

U​

 

 

 

 

 

 

Waveform​

 

 

 

 

 

 

 

 

Measurements”​

 

 

 

 

TID 3505 Atrial Pressure Measurement​

The Atrial Pressure Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic​ site, and containing a-wave, v-wave, and mean measurements. This implies that the name of the measurement is effectively post-​ coordinated from the measurement name, the Hemodynamic Measurement Group container (procedure phase) name, and the ac-​ quisition context (finding site) value.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

- Standard -​

Page 222​ DICOM PS3.16 2020a - Content Mapping Resource​

Table TID 3505. Atrial Pressure Measurement​

 

NL​

Rel with​

VT​

Concept Name​

VM​

Req​ Condition​

Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

 

1​

 

 

CONTAINER​EV(122121,DCM,"Atrial​ 1​

M​

 

 

 

 

 

 

pressuremeasurements")​

 

 

 

 

2​

>​

 

INCLUDE​

DTID 3530​

1​

M​

$LocationName=EV(363698007,​

 

 

 

 

“Hemodynamic​

 

 

SCT, "Finding Site")​

 

 

 

 

Acquisition Context”​

 

 

$LocationValue = DCID 3608​

 

 

 

 

 

 

 

 

 

 

 

 

 

 

“Atrial Source Locations”​

3​

>​

CONTAINS​INCLUDE​

DTID300“Measurement”​ 1​

M​

$Measurement = EV (109016,​

 

 

 

 

 

 

 

DCM, "A-wave peak pressure")​

 

 

 

 

 

 

 

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

Units”​

4​

>​

CONTAINS​INCLUDE​

DTID300“Measurement”​ 1​

M​

$Measurement = EV (109034,​

 

 

 

 

 

 

 

DCM, "V-wave peak pressure")​

 

 

 

 

 

 

 

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

Units”​

5​

>​

CONTAINS​INCLUDE​

DTID300“Measurement”​ 1​

M​

$Measurement = EV (6797001,​

 

 

 

 

 

 

 

SCT, "Mean blood pressure")​

 

 

 

 

 

 

 

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

Units”​

6​

>​

CONTAINS​INCLUDE​

DTID 3550 “Pressure​

1-n​

U​

 

 

 

 

 

 

Waveform​

 

 

 

 

 

 

 

 

Measurements”​

 

 

 

 

TID 3506 Venous Pressure Measurement​

The Venous Pressure Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic​ site, and containing a mean measurement. This implies that the name of the measurement is effectively post-coordinated from the​ measurement name, the Hemodynamic Measurement Group container (procedure phase) name, and the acquisition context (finding​ site) value.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

 

Table TID 3506. Venous Pressure Measurement​

NL​

Rel with​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

Parent​

 

 

 

Type​

 

1​

 

CONTAINER​EV (31724009, SCT,​

1​

M​

 

 

 

 

"Venous pressure​

 

 

 

 

 

 

measurements")​

 

 

 

2​ >​

 

INCLUDE​

DTID 3530 “Hemodynamic​

1​

M​

$LocationName = EV​

 

 

 

Acquisition Context”​

 

 

(363698007, SCT, "Finding​

 

 

 

 

 

 

Site")​

$LocationValue = DCID 3607​

“Venous Source Locations”​

- Standard -​

 

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 223​

 

NL​

Rel with​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

3​

>​

CONTAINS​INCLUDE​

DTID 300 “Measurement”​

1​

M​

$Measurement=EV(6797001,​

 

 

 

 

 

 

 

SCT, "Mean blood pressure")​

 

 

 

 

 

 

 

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

Units”​

4​

>​

CONTAINS​INCLUDE​

DTID 3550 “Pressure​

1-n​

U​

 

 

 

 

 

Waveform Measurements”​

 

 

 

TID 3507 Ventricular Pressure Measurement​

The Ventricular Pressure Measurement Template consists of a CONTAINER, with an acquisition context of the measurement ana-​ tomic site, and containing systolic and end-diastolic measurements. This implies that the name of the measurement is effectively​ post-coordinated from the measurement name, the Hemodynamic Measurement Group container (procedure phase) name, and the​ acquisition context (finding site) value.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

 

Table TID 3507. Ventricular Pressure Measurement​

 

NL​ Rel with​

VT​

Concept Name​ VM​ Req​

Condition​

Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

1​

 

 

CONTAINER​EV (122122, DCM,​ 1​

M​

 

 

 

 

 

 

"Ventricular​

 

 

 

 

 

 

 

 

pressure​

 

 

 

 

 

 

 

 

measurements")​

 

 

 

 

2​

>​

 

INCLUDE​

DTID 3530​

1​

M​

 

$LocationName = EV​

 

 

 

 

“Hemodynamic​

 

 

 

(363698007, SCT, "Finding​

 

 

 

 

AcquisitionContext”​

 

 

 

Site")​

 

 

 

 

 

 

 

 

$LocationValue = DCID 3609​

 

 

 

 

 

 

 

 

“Ventricular Source Locations”​

3​

>​

CONTAINS​INCLUDE​

DTID 300​

1​

MC​IFF $LocationValue​

$Measurement = EV​

 

 

 

 

“Measurement”​

 

 

selected in row 2 is​

(276780008, SCT, "Left​

 

 

 

 

 

 

 

(87878005, SCT, "Left​

Ventricular Systolic blood​

 

 

 

 

 

 

 

Ventricle") or subsite​

pressure")​

 

 

 

 

 

 

 

thereof​

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Units”​

4​

>​

CONTAINS​INCLUDE​

DTID 300​

1​

MC​IFF $LocationValue​

$Measurement = EV​

 

 

 

 

“Measurement”​

 

 

selected in row 2 is​

(276781007, SCT, "Left​

 

 

 

 

 

 

 

(87878005, SCT, "Left​

Ventricular End Diastolic​

 

 

 

 

 

 

 

Ventricle") or subsite​

pressure")​

 

 

 

 

 

 

 

thereof​

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Units”​

5​

>​

CONTAINS​INCLUDE​

DTID 300​

1​

MC​IFF $LocationValue​

$Measurement = EV​

 

 

 

 

“Measurement”​

 

 

selected in row 2 is​

(276772001, SCT, "Right​

 

 

 

 

 

 

 

(53085002, SCT, "Right​Ventricular Systolic blood​

 

 

 

 

 

 

 

Ventricle") or subsite​

pressure")​

 

 

 

 

 

 

 

thereof​

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Units”​

- Standard -​

Page 224​

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

NL​ Rel with​

VT​

Concept Name​ VM​ Req​

Condition​

Value Set Constraint​

Parent​

 

 

Type​

 

6​ >​ CONTAINS​INCLUDE​

DTID 300​

1​ MC​IFF $LocationValue​

$Measurement = EV​

 

 

“Measurement”​

 

selected in row 2 is​

(276774000, SCT, "Right​

 

 

 

 

(53085002, SCT, "Right​Ventricular End Diastolic​

 

 

 

 

Ventricle") or subsite​

pressure")​

 

 

 

 

thereof​

 

7​ >​ CONTAINS​INCLUDE​ DTID 300​

1​ MC​IFF $LocationValue​

“Measurement”​

selected in row 2 is​

 

(45503006, SCT,​

 

"Common Ventricle")​

$Units = DCID 3500 “Pressure​

Units”​

$Measurement = EV (122194,​ DCM,"VentricularSystolicblood​ pressure")​

$Units = DCID 3500 “Pressure​

Units”​

8​ >​ CONTAINS​INCLUDE​ DTID 300​

1​ MC​IFF $LocationValue​

“Measurement”​

selected in row 2 is​

 

(45503006, SCT,​

 

"Common Ventricle")​

$Measurement = EV (122191,​ DCM, "Ventricular End Diastolic​ pressure")​

$Units = DCID 3500 “Pressure​

Units”​

9​ >​ CONTAINS​INCLUDE​ DTID 3550​

1-n​ U​

“Pressure​

 

Waveform​

 

Measurements”​

 

TID 3508 Gradient Measurement​

The Gradient Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic site, and​ containing the gradient measurement. This implies that the name of the measurement is effectively post-coordinated from the meas-​ urement name, the Hemodynamic Measurement Group container (procedure phase) name, and the acquisition context (finding site)​ value.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

 

Table TID 3508. Gradient Measurement​

 

NL​ Rel with​

VT​

Concept Name​

VM​

Req​ Condition​

Value Set Constraint​

 

Parent​

 

 

 

Type​

 

1​

 

CONTAINER​EV (122123, DCM,​

1​

M​

 

 

 

 

"Gradientassessment")​

 

 

 

2​

>​

INCLUDE​

DTID 3530​

1​ MC​ XOR with Rows 3​$LocationName = EV​

 

 

 

“Hemodynamic​

 

& 4​

(363698007,SCT,"FindingSite")​

 

 

 

Acquisition Context”​

 

IFFsinglelocation​$LocationValue = DCID 3610​

 

 

 

 

 

 

 

 

 

 

is appropriate​

“Gradient Source Locations”​

3​

>​

INCLUDE​

DTID 3530​

1​ MC​ XOR with Row 2​$LocationName = EV (121116,​

 

 

 

“Hemodynamic​

 

 

DCM, "Proximal Finding Site")​

 

 

 

Acquisition Context”​

 

IFFaduallocation​

 

 

 

 

 

is appropriate​

$LocationValue = DCID 3630​

 

 

 

 

 

 

“Cardiovascular Anatomic​

 

 

 

 

 

 

Locations”​

- Standard -​

 

 

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 225​

 

NL​ Rel with​

VT​

Concept Name​

VM​

Req​ Condition​

Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

4​

>​

 

INCLUDE​

DTID 3530​

1​ MC​ XOR with Row 2​$LocationName = EV (121117,​

 

 

 

 

“Hemodynamic​

 

 

DCM, "Distal Finding Site")​

 

 

 

 

Acquisition Context”​

 

IFFaduallocation​

 

 

 

 

 

 

is appropriate​

$LocationValue = DCID 3630​

 

 

 

 

 

 

 

“Cardiovascular Anatomic​

 

 

 

 

 

 

 

Locations”​

5​

>​

CONTAINS​INCLUDE​

DTID 300​

1-n​

M​

$Measurement=EV(251081004,​

 

 

 

 

“Measurement”​

 

 

SCT, "Pressure Gradient")​

 

 

 

 

 

 

 

$Units = DCID 3500 “Pressure​

 

 

 

 

 

 

 

Units”​

 

 

 

 

 

 

 

$Derivation = DCID 3627​

 

 

 

 

 

 

 

“Measurement Type”​

6​

>​

CONTAINS​INCLUDE​

DTID 3550 “Pressure​

1-n​

U​

 

 

 

 

 

Waveform​

 

 

 

 

 

 

 

Measurements”​

 

 

 

Content Item Descriptions​

Row 5​ Is used to encode the gradient measurements. Uses TID 300 “Measurement” for the Content Item structure of the​ measurements. That Template allows an INFERRED FROM relationship to other numeric measurements used in​ the computation or derivation of the current measurement. In the case of a gradient measurement, the pressure​ measurements at the distal and proximal locations may thus be explicitly conveyed.​

TID 3509 Blood Velocity Measurement​

The Blood Velocity Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic​ site, and containing primary (not derived) velocity measurements, e.g., from a Doppler catheter. Derived velocity measurements may​ be recorded using TID 3560 “Derived Hemodynamic Measurements”.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

 

Table TID 3509. Blood Velocity Measurement​

 

NL​

Rel with​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

1​

 

 

CONTAINER​EV (122124, DCM, "Blood​

1​

M​

 

 

 

 

 

velocity measurements")​

 

 

 

2​

>​

 

INCLUDE​

DTID 3530 “Hemodynamic​

1​

M​

$LocationName = EV​

 

 

 

 

Acquisition Context”​

 

 

(363704007, SCT, "Procedure​

 

 

 

 

 

 

 

site")​

 

 

 

 

 

 

 

$LocationValue = BCID 3606​

 

 

 

 

 

 

 

“Arterial Source Locations”​

3​

>​

CONTAINS​INCLUDE​

DTID 300 “Measurement”​

1-n​

M​

$Measurement = DCID 3612​

 

 

 

 

 

 

 

“BloodVelocityMeasurements”​

 

 

 

 

 

 

 

$Units = EV (mm/s, UCUM,​

 

 

 

 

 

 

 

"mm/s")​

4​

>​

CONTAINS​INCLUDE​

DTID 3550 “Pressure​

1-n​

U​

 

 

 

 

 

Waveform Measurements”​

 

 

 

- Standard -​