If an acquisition is a repeat because an earlier acquisition was unsatisfactory, this may be recorded along with |
|
a coded reason. This is intended to help with subsequent analysis by providing a priori information about why |
|
the study might be flagged as an outlier with higher dose exposure values than usual for the type of study. |
Row 6f |
The DateTime that the application of X-Rays started for this irradiation event. This shall correspond to the start |
|
of the first irradiation in the Irradiation Event, which defines the starting point for the calculation of the contents |
|
of the Row 21 "CT Dose" CONTAINER. |
|
Row 8 |
Total time the patient has received X-Ray exposure during the irradiation event. |
|
Row 10 |
Thevalueofthenominalwidth(referencedtothelocationoftheisocenteralongthezaxis)ofasinglecollimated |
|
slice in mm. |
|
Row 11 |
The value of the nominal width (referenced to the location of the isocenter along the z axis) of the nominal total |
|
collimation in mm over the area of active X-Ray detection (z-coverage). |
|
Row 12 |
Pitch Factor: For Spiral Acquisition, the Pitch Factor is the ratio of the Table Feed per Rotation to the Nominal |
|
Total Collimation Width. For Sequenced Acquisition, the Pitch Factor is the ratio of the Table Feed per single |
|
sequenced scan to the Nominal Total Collimation Width. |
|
Row 14 |
CT X-Ray source parameters related to the acquisition. For each X-Ray source an item must be present. For |
|
multi-energy acquisitions, multiple items may be present for each X-Ray source, each item describing one |
|
energy level. |
|
Row 15 |
Identification of the X-Ray source. Identifies the particular X-Ray source (in a multi-source CT system) for |
|
which the set of X-Ray source parameter values is reported. |
|
Row 16 |
KVP value as measured/recorded by system. |
|
Row 18 |
Mean tube current as measured/recorded by system. |
|
Row 19 |
Exposure time as measured/recorded by the system per rotation. |
|
Row 20 |
Thickness of an equivalent filter constructed from aluminum, in case of multi-source CT systems AND if Row |
|
4 is not present |
|
Row 21 |
CT Dose for one acquisition |
|
Row 22 |
"Mean CTDIvol" refers to the average value of the CTDIvol applied within this acquisition. |
|
|
CTDIvol is the volume CTDIw, where CTDIw is the weighted computed tomography dose index 100 as defined |
|
in IEC 60601-2-44. |
|
|
For Sequenced and Spiral scanning, CTDIvol = CTDIw / Pitch Factor. |
|
|
ForStationaryandFreescanning,CTDIvol =CTDIw CumulativeExposureTime/ExposureTimePerRotation. |
|
According to IEC 60601-2-44 Ed 3 for Constant Angle Acquisition may be calculated as CTDIvol = (CTDIw / |
|
Current Time Product (mAs)) X-Ray Tube Current (mA) x (Nominal Total Collimation Width (mm) / Table |
|
Speed (mm/s)). |
|
|
Note |
|
|
TheratioCTDIw/CurrentTimeProductisevaluatedindependentlyoftheConstantAngleAcquisition |
|
but with the same settings of tube voltage and Total Collimation Width as those of the Constant Angle |
|
Acquisition. |
|
|
See also CTDIvol (0018,9345) and Spiral Pitch Factor (0018,9311) in the “Enhanced Computed Tomography |
|
Image IOD” in PS3.3. |
|
Row 23 |
The type of phantom used for CTDI measurement according to IEC 60601-2-44(e.g., Head 16 cm diameter |
|
PMMA, Body 32 cm diameter PMMA). |
|
Row 24 |
The CTDIfree air Calculation Factor is the CTDIfree air per mAs, expressed in units of mGy/mAs. The CTDIfree air |
|
Calculation Factor may be used in one method calculating Dose. For example, for this acquisition, Effective |