DICOM PS3.17 2020a - Explanatory Information |
Page 705 |
Table |
|
Fixed Point |
Isocenter |
|
X-Ray
Table
3D pre-intervention
Table
Fixed Point
Isocenter
X-Ray
Table
3D post-intervention
Figure TTT.2.6-2. One frame of two 3D reconstructions at two different table positions
The volume is positioned in the spatial coordinates identified by the frame of reference, which is common to the two volumes. Therefore, the position of the slices of both volumes is defined with respect to the same patient origin.
TTT.2.6.3.1.5 Plane Orientation (Patient) Macro Recommendations
The slices can be oriented in any relation wrt. the patient coordinate system. The plane orientation is expected to be the same for the two volumes; however it could be different without compromising the registration.
TTT.2.6.4 Example
In this example, two rotational images are acquired; the first one before the intervention and the second one after the intervention. They are encoded with the Instance UIDs "C1" and "C2" respectively.
In both rotational acquisitions, the patient position with respect to the table is head-first prone, and the table is not rotated nor tilted with respect to the Isocenter. The patient coordinates and the Isocenter coordinates are then aligned on x, y and z.
The patient origin is defined by the application as a fixed point on the table.
During the first rotational acquisition, the table position with respect to the Isocenter in the lateral direction [x] is +20mm, in the vertical direction [y] is +40mm, and in the longitudinal direction [z] is +60mm.
During the second rotational acquisition, the table position with respect to the Isocenter in the lateral direction [x] is -10mm, in the vertical direction [y] is +80mm, and in the longitudinal direction [z] is +110mm.
Thesecondacquisitionisperformedwitharelativetablemovementof(-30,40,50)mmvs.thefirstacquisitioninthepatientcoordinates system. Therefore, for a given 3D slice "i" of the two volumes, the Image Position (Patient) (0020,0032) of the second volume is translated of (+30,-40,-50) mm vs. the Image Position (Patient) (0020,0032) of the first volume.
The two reconstructions are performed with the same number of rows, columns and slices, and both at the same resolution of 0.2 mm/voxel.Notethatiftheresolutionwasdifferent,theImagePosition(Patient)(0020,0032)ofthesecondvolumewouldbeadditionally translated by the shift of the TLHC pixels relative to the center of the volume, because both volumes are centered at the Isocenter.
The reconstructions are encoded in two X-Ray 3D Angiographic instances of Instance UIDs "Z1" and "Z2" respectively.